Sartorius Biowelder Total Containment Specifications

General IconGeneral
Sterile ConnectionYes
Total ContainmentYes
Power Supply100-240 VAC, 50/60 Hz
Welding TechnologyThermal
Welding TimeProgrammable
Cooling TimeProgrammable
WeightVaries by model
ComplianceCE
Tube Size Range1/4" to 1" OD (6.35 mm to 25.4 mm)

Summary

Safety and Hazard Information

Hand Injury, Emergency Stop, Power Disconnection

Warnings on internal access, emergency stop, and power cord removal for safety.

ESD Precautions and Grounding

Guidelines for handling ESD-sensitive devices, including grounding straps and mats.

Critical Section Welding Safety and Damage Handling

Ensuring safe welding, avoiding critical section discontinuity, and handling device damage.

High Temperature Surface Hazard Warning

Warning about hot surfaces and the need for protective measures to prevent burns.

Operating Environment and Power Supply Safety

Covers voltage, fluid protection, temperature, humidity, UPS, and network configuration safety.

Component Usage and Network Security

Guidelines on using blades, cables, fuses, and ensuring network connection security.

User Login and Auto-Login Security Risks

Importance of logging out and risks of auto-login for Supervisor accounts.

Document Conventions and General Information

Document Scope and Symbol Meanings

Defines document audience, browser used, and explains CAUTION/Information symbols.

Warranty, Liability, and Proper Use

Details warranty coverage, liability, and defines proper device usage.

Patent and Manufacturer Information

Mentions trademark registration and manufacturer's rights to make changes.

Introduction to Biowelder® TC

Welding Technology and Device Features

Explains thermal welding, sterile process, compatible tubes, and key device features.

Safety Precautions and Device Limitations

Highlights safety measures during operation and outlines device limitations.

Functionality and Welding Process Steps

Details the step-by-step welding process, automated functions, and data logging.

Cleaning, Blocked Devices, and Decontamination

Provides guidelines for cleaning, handling blocked devices, and decontamination procedures.

Device Start-up Procedures

Delivery Kit and Component Familiarization

Lists delivery kit contents and identifies key device components via diagram.

Installation, Login, and Axis Initialization

Guides on installation, power connection, user login, and axis initialization.

Blade Handling and Welding Process Initiation

Step-by-step instructions for blade insertion and initiating the welding process.

Completing the Welding Sequence

Steps after welding: blade removal, tube removal, quality check, and waste disposal.

Device Operation and User Interfaces

User Interfaces and Mutual Exclusion Rules

Describes touch screen/Web browser operation and rules for mutual exclusion.

Touch Screen Navigation and Controls

Illustrates menu structure, symbols, and operator controls for the touch screen.

User Login/Logout and Date/Time Settings

Details user login/logout and setting date/time on the touch screen.

Language, Tube Type, and Network Reset

Guides on changing language, tube type, and resetting network parameters.

User Management and Log Data Access

Covers user management functions and accessing log data via touch screen/web browser.

Web Browser Connection and Settings

Instructions for connecting via web browser and configuring general network settings.

Web Browser User Management and Info

Covers user management and info menu via web browser.

Automatic Logout Functionality

Describes the automatic logout feature, exceptions, and interaction with auto-login.

Sequences and Troubleshooting

Sequence Execution and Critical Section Risks

Explains sequences like Initialize Axes, Weld, and the critical section risks.

Aborting Welds, Rewelding, and Reverse

Details handling critical section aborts, rewelding, and the reverse sequence.

Temperature Validation Procedure

Guides on validating the built-in temperature sensor using external measurement.

Common Operational Problems

Addresses issues like device blockages, incorrect tube insertion, and memory card errors.

Network and Password Recovery

Covers network configuration, web browser settings, and password recovery.

Appendix and Technical Data

Copyright, Disposal, and Technical Specifications

Details copyright, disposal, and lists key technical specifications of the device.

Technical Contact and Device Dimensions

Provides contact information and illustrates external dimensions of the Biowelder® TC.

User Groups, Log Content, and Tube Holders

Explains user groups, log content, and tube holder identification.

Error Reports and Warning Messages

Lists common error reports and warning messages with their causes and remedies.