46 Operating Manual Biowelder
®
Total Containment
Appendix
Warning messages Cause Elimination
There are no valid parameters on this
machine. Please load parameters.
There are no parameters on
the device.
Switch machine on and off.
If there is still no current data,
contact your Sartorius Stedim
Biotech Service Representative.
SD missing. Please insert SD and
retry!
No SD card inserted Insert card in SD slot and repeat
the action again.
SD not valid Wrong SD card inserted Use the SD card supplied by
Sartorius Stedim Biotech.
Login failed Wrong user name or password
entered
Enter correct user name with
the associated password
Insert validation holder 1. Validation holder not inserted
2. Validation holder inserted,
identification defective
1. Insert validation holder
2. Contact Sartorius Stedim
Biotech Service
Representative.
Remove validation holder 1. Validation holder was not removed
2. Validation holder was removed,
identification defective
1. Remove validation holder
2. Contact your Sartorius
Stedim Biotech Service
Representative.
Run stopped by user User interrupts a sequence Confirm message with OK
Log transfer stopped by user Abort by the user while data being
written to the SD card
Confirm message with OK