Operating Instructions Cubis
®
MCA 111
Technical Data
15.7.2 Models MCA524S | MCA524P | MCA324S | MCA324P
MCA524S MCA524P MCA324S MCA324P
Unit Value Value Value Value
Scale interval (d) mg 0.1 0.1/0.2/0.5 0.1 0.1/0.2/0.5
Maximum capacity (Max) g 520 120/240/
520
320 80/160/320
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.08 0.08 0.08 0.08
Standard deviation of the load values,
typical value
mg 0.04 0.04 0.04 0.04
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.1 0.15 0.1 0.1
Standard deviation of the load values,
typical value
mg 0.05 0.05 0.05 0.05
Linearity deviation
Tolerance mg 0.4 0.5 0.3 0.5
Typical value mg 0.2 0.2 0.2 0.2
Deviation when load is off-center,
positions according to OIML R76
Test weight g 200 200 200 200
Tolerance mg 0.3 0.4 0.3 0.4
Typical value mg 0.2 0.2 0.2 0.2
Sensitivity drift between +10°C and +30°C ppK 1 1 1 1
Tare maximum capacity:
Less than 100% ofmaximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according
toDirective2014/31/EU
mg 10 10 10 10
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 82 82 82 82
Typical minimum initial weighing mg 82 82 82 82
Typical stabilization time s 1 1 1 1
Typical measurement time s 3 3 3 3