Operating Instructions Cubis
®
 MCA  79
    System Settings 
7.7.20  Parameters in the “Device Settings / Preload” Menu
Parameter Setting Values Explanation
Preload Set preload  − Sets the load present on the device as a preload and 
therefore changes the zero point of the device.
 − Reduces the maximum capacity by the set preload.
Delete preload  − Deletes the set preload.
 − Changes the zero point and maximum capacity back to the 
values they were before the preload was set.
7.7.21  Parameters in the “Device Settings / Calibration / Adjustment Report” Menu
Parameter Setting Values Explanation
Automatic printout 
for isoCAL
On Activates the automatic printout for the “isoCAL” function.
Off* Deactivates the automatic printout for the “isoCAL” function.
Automatic printout 
for cal./adj. task
On Activates the automatic printout for the functions of the 
“calibration/adjustment” tasks.
Off* Deactivates the automatic printout for the functions 
ofthe“calibration/adjustment” tasks.
Print profile List of print profiles Selection of a print profile via which the printout must 
becarriedout.
* Factory setting
7.7.22  Parameters in the “Device Settings / minUSP” Menu
Parameter Setting Values Explanation
minUSP function Off* If this extension has been activated in the QAPP Center: 
Deactivates the minimum initial weight.
On If this extension has been activated in the QAPP Center: 
Activates the minimum initial weight.
Starting point of 
operating range
User input If this extension has been activated in the QAPP Center: 
Definethe starting point value.
Mark weight values 
< minUSP as invalid
Off* If this extension has been activated in the QAPP Center: 
Deactivates the identification of weight values that are less than 
the minimum initial weight.
On If this extension has been activated in the QAPP Center: 
Activates the identification of weight values that are less than 
the minimum initial weight and prevents them being adopted. 
These values are not only marked, they cannot be adopted.
* Factory setting