116  Operating Instructions Cubis
®
 MCA
  Technical Data
15.9  isoCAL Function
15.9.1  Models MCA225S | MCA225P | MCA125S | MCA125P | MCA324S | MCA324P | 
MCA224S | MCA124S | MCA2203S | MCA2203P | MCA1203S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling
15.9.2  Models MCA524S | MCA524P | MCA5203S | MCA5203P | MCA3203S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 1.5
After a time interval h 6
After successful leveling
15.9.3  Models MCA623S | MCA623P | MCA323S | MCA5202S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 2
After a time interval h 12
After successful leveling
15.10  Data Storage Device
Maximum number of datasets 150,000