82 Operating Instructions Cubis
®
MCA
System Settings
7.7.28 Parameters in the “Device Settings / Start-up Behavior” Menu
Parameter Setting Values Explanation
Initial zeroing /
taring
Off Deactivates automatic zeroing and taring when starting
thedevice.
On* Activates automatic zeroing and taring when starting the device.
Automatic login Off Deactivates “Automatic login”. No user is logged in
automatically when starting the device.
Administrator* Activates “Automatic login”. The administrator is logged in
automatically when starting the device.
Operator Activates “Automatic login”. The selected user is logged in
automatically when starting the device.
Automatic start of
last task
Off* No task is started automatically after device login.
On After logging-in on the device, the task being performed
bythelast user to be logged in starts automatically.
* Factory setting
7.7.29 Parameters in the “Device Settings / Display Properties” Menu
Parameter Setting Values Explanation
Display brightness Bright Sets the illumination on the operating display to “bright”.
Medium Sets the illumination on the operating display to “medium”.
ECO mode* Activates eco mode. For longer periods of downtime,
theoperating display darkens automatically.
On/off switch Standby mode* The [Standby] button switches the device into standby mode.
The device can be reactivated by pressing the [Standby] button.
Off The [Standby] button switches the device off. The device must
be reactivated by pressing the ON key.
Color scheme Sartorius Standard* If the QAPP extension has been activated: Selects the color
scheme for the “Sartorius Standard” operating display.
Additional color schemes can be unlocked via the QAPP Center.
* Factory setting