Loading...
ID code characters I
1)
Stat Status
ID Identifier
L ID Weighing series no.
W ID Weight set number
Nom. Exact calibration weight
S ID Sample ID
NUM Numeric input
T1 Application tare memory 1
N Net weight (T1 = 0)
N1 Net weight (T1# 0)
Qnt Quantity
Prc Percentage
nRef Reference sample quantity
pRef Reference percentage
wRef Average piece weight
Wxx% Reference percentage weight
mDef Target value for animal
weighing
Mul Multiplication factor for
animal weighing
x-Net Result in animal weighing
x-Res Calculated result in
animal weighing
Res Result using equation
(Calculation)
Setp Target value for
checkweighing
Min Lower limit for
checkweighing
Max Upper limit for
checkweighing
Time Time that a value was stored
Compxx No. of components
in formulation
Tot.cp Total weight in formulation
n Transaction counter
Total Sum of all values
Avg Average in statistics
s Standard deviation
srel Variation coefficient
Diff Difference between
maximum and minimum
103
Data output example: + 1255.7 g
Position 12345678910111213141516
+* * *1255 . 7*g* *CRLF
Position 1: Plus or minus sign or space
Position 2: Space
Position 3–10: Weight with a decimal point; leading zeros = space
Position 11: Space
Position 12–14: Unit symbol or space
Position 15: Carriage return
Position 16: Line feed
Data Output With ID Code
When data with an ID code is output, the ID code consisting of 6 characters precedes the
data with the 16-character format. These 6 characters identify the subsequent value.
12345678910111213141516171819202122
KIIIIII+*DDDDDDDD * UUUCRLF
***** ........ ***
* ********
I: ID code character
1)
U: Unit symbol
1)
see “Toggle
*: Space between Weight Units”
D: Digit or letter CR: Carriage return
LF: Line feed
1
) depends on scale type; e.g., not all units and characters are available on scales verified
for use in legal metrology
Special Codes
12345678910111213141516171819202122
Stat****** * *––******CRLF
HH
LL
C
*: Space L: Underload
– –: Weight L L: Underload in checkweighing
H: Overload C: Calibration/adjustment
H H: Overload in checkweighing
Error Codes
12345678910111213141516171819202122
Stat*****Er r *###****CRLF
*: Space # # #: Error code number