Model-Specific Specifications:
Model FC12CCE-SXCE FC6CCE-SXCE FC12CCE-IXCE
Type isoTEST in conjunction with BD BF
Accuracy class * KKK
Scale interval d* g 0.1 0.1 0.5
Max. weighing capacity* g 12,000 6,200 12,000
Verification scale interval e* g 1 1 0.5
Min. capacity* g 5 5 25
Max. capacity kg 50 50 50
Tare range (subtractive) <100% of the max. weighing capacity
Electronically compensated preload g 1,200 1,240 1,200
(without restricting weighing range)
Max. preload when starting g 8,200 2,440 8,200
calibration/adjustment
(scale must be zeroed)
Application range according to CD* g 5–12,000 5–6,200 25–12,000
Response time (average) s 1 1 1
Allowable ambient operating +10° to +30°C (50° to 86°F)
temperature
Pan size mm 218+ 200 218+ 200 218+200
Net weight, approx. kg 7.3 7.3 7.3
Dust and water protection rating IP54
according to EN 60529
Model FC34EDE-HXCE, FC16EDE-HXCE, FC34EDE-PXCE, FC12EDE-PXCE, FC64EDE-SXCE,
FCA34EDE-HXCE FCA16EDE-HXCE FCA34EDE-PXCE FCA12EDE-PXCE FCA64EDE-SXCE
Type isoTEST in conjunction with BF BF
Accuracy class * KKKKK
Scale interval d* g 0.1 0.1 0.1/0.2/0.5 0.1/0.2 1
Max. weighing capacity* kg 34 16 8/16/34 6/12 64
Verification scale interval e* g 111110
Min. capacity* g 555550
Tare range (subtractive) <100% of the max. weighing capacity
Max. capacity kg 130
Electronically compensated preload kg 444413
(without restricting weighing range)
Max. preload when starting kg ca. 21 ca. 19 ca. 21 ca. 10 ca. 45
calibration/adjustment
(scale does not have to be zeroed)
Application range according to CD* g 5–34,000 5–16,000 5–34,000 5–12,000 50–64,000
Response time (average) s 1.5
Allowable ambient operating +10° to +30°C (50° to 86°F)
temperature
Pan size mm 300+ 400
Net weight, approx. kg 16.0
Dust and water protection rating IP65
according to EN 60529