126
Specifications
Model FCA150IGG-HXCE FCA300IGG-HXCE
Type isoTEST in conjunction with HC BF
Accuracy class * KK
Scale interval d* g 1 20
Max. weighing capacity* kg 150 300
Verification scale interval e* g 10 20
Min. capacity* g 50 1000
Tare range (subtractive) <100% of the max. weighing capacity
Max. overload capacity kg 600 600
Tare range (subtractive) kg 150 300
Electronically compensated preload kg 30 60
(without restricting weighing range)
Max. preload when starting kg approx. 120 approx. 300
calibration/adjustment (scale display
does not have to be zeroed)
Application range according to CD* kg 0.05-150 1-300
Response time (average) s 1.5 1.5
Allowable ambient operating 0 to +40°C (273 to 313°K; +32° to 104°F)
temperature
External standard calibration weight kg 50 (F2) 50 (F1)
(of at least accuracy class...)
Platform dimensions mm 800 + 600
Net weight, approx. kg 70
Dust and water protection rating IP67
according to EN 60529
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area