137
Models Verified by the Manufacturer, with EC Type Approval
General Specifications
AC power source/power requirements AC adapter, 230 or 115 V, +15% ... – 20%
Frequency 48 – 60 Hz
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Power consumption 16 VA: maximum; 9 VA: average
Hours of operation with fully charged
YRB 06 Z external battery pack, approx. 14 h
Selectable application programs Mass unit conversion, counting, weighing in percent, animal weighing, recalculation,
calculation, density determination, differential weighing, checkweighing, time-controlled
functions, totalizing, formulating, statistics, 2nd tare memory, IDs, product data memories
Built-in interface RS-232 C
Format: 7-bit ASCII, 1 start bit,
1 or 2 stop bits
Parity: odd, even or space
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Specifications of the Individual Models:
Model LA310S-0CE LA230S-0CE LA230P-0CE LA120S-0CE
Type isoTEST in conjunction with BC BF
Accuracy class* kkkk
Scale interval, d* mg 0.1 0.1 0.1/0.2/0.5 0.1
Maximum weighing capacity, Max* g 310 230 60/120/230 120
Verification scale interval, e* g 0.001
Minimum capacity, Min* g 0.01
Tare range (subtractive) ≤ 100% of the maximum capacity
Application range according to CD* g 0.01 – 310 0.01 – 230 0.01 – 230 0.01 – 120
Response time (average) s 2
Allowable operating temperature 273 ... 313 K (0 ... +40 °C, 32 °F ... 104 °F) with isoCAL function
Selectable weight units Grams, milligrams
External calibration weight value g 200 + 100 (E2) 200 (E2) 200 (E2) 100 (E2)
(of at least accuracy class...)
Other permissible external calibration g 200 (E2) 100, 150 (E2) 100, 150 (E2) 50 (E2)
weights (of at least accuracy class...)
Pan size mm d 90
Dimensions (W + D + H) mm 261 + 381 + 361
Net weight approx. kg 8.7
Dust and water protection rating
according to EN 60529
1
) IP42
1
) = Specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories.”
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area