138
Model LA1200S-0CE LA620S-0CE LA220S-0CE LA620P-0CE
Type isoTEST in conjunction with BD BF
Accuracy class* kKKK
Scale interval, d* g 0.001 0.001 0.001 0.001/0.002/0.005
Maximum weighing capacity, Max* g 1,200 620 220 120/240/620
Verification scale interval, e* g 0.01 0.01 0.01 0.01
Minimum capacity, Min* g 0.1 0.02 0.02 0.02
Tare range (subtractive) 100% of the max. weighing capacity
Application range according to CD* g 0.1 – 1,200 0.02 – 620 0.02 – 220 0.02 – 620
Response time (average) s 1.5
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams, kilograms
External calibration weight value g 1,000 (E2)
(of at least accuracy class...)
Pan size mm d 130
Dimensions (W + D + H) mm 261 + 381 + 147
Net weight, approx. kg 8.3 6.9 6.9 6.9
Dust and water protection rating
according to EN 60529
2
) IP54
Model LA8200S-0CE LA8200P-0CE LA6200S-0CE LA4200S-0CE LA2200S-0CE
Type isoTEST in conjunction with BD BF
Accuracy class* KKKKK
Scale interval, d* g 0.01 0.01/0.02/0.05 0.01 0.01 0.01
Maximum weighing capacity, Max* g 8,200 2,000/4,000/8,200 6,200 4,200 2,200
Verification scale interval, e* g 0.1 0.1 0.1 0.1 0.1
Minimum capacity, Min* g 0.5 0.5 0.5 0.5 0.5
Tare range (subtractive) 100% of the max. weighing capacity
Application range according to CD* g 0.5 – 8,200 0.5 – 8,200 0.5 – 6,200 0.5 – 4,200 0.5 – 2,200
Response time (average) s 2 2 1.5 1.5 1.5
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams, kilograms
Pan size mm 218 + 200
Dimensions (W + D + H) mm 261 + 381 + 86
Net weight. approx. kg 6.5
Dust and water protection rating
according to EN 60529
2
) IP54
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified
by the Sartorius Service Center): For balances of accuracy class k: +15°C to +25°C (+59°F to +77°F)
For balances of accuracy class K: +10°C to +30°C (+50°F to +86°F)
2
) = Specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories.”
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area