139
Model LA820-0CE LA2200P-0CE LA5200P-0CE LA12000S-0CE LA6200-0CE LA2200-0CE
Type isoTEST in conjunction with BD BF
Accuracy class* KKKK KK
Scale interval, d* g 0.01 0.01/0.02/ 0.01/0.02/ 0.1 0.1 0.1
0.05 0.05/0.1
Maximum weighing capacity, Max* g 820 400/800/ 1,200/2,400/ 12,000 6,200 2,200
2,200 3,800/5,200
Verification scale interval, e* g 0.1 0.1 0.1 1 1 0.1
Minimum capacity, Min* g 0.5 0.5 0.5 5 5 5
Tare range (subtractive) ≤ 100% of the max. weighing capacity
Application range according to CD* g 0.5 – 820 0.5 – 2,200 0.5 – 5,200 5 – 12,000 5 – 6,200 5 – 2,200
Response time (average) s 1.5 1.5 1.5 1 1 1
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams, kilograms
Pan size mm 218 + 200
Dimensions (W + D + H) mm 261 + 381 + 86
Net weight. approx. kg 6.5
Dust and water protection rating
according to EN 60529
2
) IP54
Model LA12000P-0CE LA34001S-0CE LA16001S-0CE LA34001P-0CE LA34000-0CE
Type isoTEST in isoTEST in isoTEST in isoTEST in isoTEST in
conjunction conjunction conjunction conjunction conjunction
with BD BF with BF BF with BF BF with BF BF with BF BF
Accuracy class* KKKKK
Scale interval, d* g 0.1/0.2/0.5 0.1 0.1 0.1/0.2/0.5 1
Maximum weighing capacity, Max* kg 3/6/12 34 16 8/16/34 34
Verification scale interval, e* g 11111
Minimum capacity, Min* g 555550
Tare range (subtractive) ≤ 100% of the max. weighing capacity
Application range acc. to CD* g 5 – 12,000 5 – 34,000 5 – 16,000 5 – 34,000 50 – 34,000
Response time (average) s 1 1.5 1.5 1.5 1
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams and Grams and Grams and Grams and Kilograms
kilograms kilograms kilograms kilograms
Pan size mm 218 + 200 300 + 400
Dimensions (W + D + H) mm 261 + 381 + 86 321 + 546 + 120 321 + 546 + 120 321 + 546 + 120 321+546+120
Net weight, approx. kg 6.5 16.0 16.0 16.0 16.0
Dust and water protection rating
According to EN 60529
2
) IP54 IP44 IP44 IP44 IP44
1
) = With the isoCAL function deactivated, the verified balances can be used within the limited temperature range (can only be modified by
the Sartorius Service Center): For balances of accuracy class k: +15°C to +25°C (+59°F to 77°F)
For balances of accuracy class K: +10°C to +30°C (+50°F to 86°F)
2
) = Specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories“
* CD = Council Directive 90/384/EWG for non-automatic weighing instruments used within the European Economic Area