EasyManuals Logo

Sartorius LA Reference Series User Manual

Sartorius LA Reference Series
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
55
Additional Functions
In addition to the functions for:
alphanumeric input,
taring (not during alphanumeric input),
printing,
you can also access the following
functions from this application:
Calibration/Adjustment
G Press the
Cal soft key
> See “Calibration/Adjustment” for
further instructions
Toggling to the Next Application
G Press d
> See the section on the corresponding
application program for further
instructions
Setup (setting parameters)
G Press o
> See “Configuration” for
further instructions
Turning Off the Balance
G Press e
> The balance shuts off
Preparation
To calculate a value in percent, the reference percentage must be known. This value can
be entered into the weighing-in-percent program in one of three ways:
The last reference percentage entered is loaded and displayed when you turn on the
balance. Place the corresponding weight on the balance and initialize the weighing-in-
percent program;
With automatic initialization switched on (Setup: App: Auto start application when power
goes on: On), the balance goes into the “weighing in percent” mode when you turn it on
and loads the last reference percentage entered as well as the corresponding reference
weight;
Enter the reference weight using the numeric keys and store it (
Wxx% soft key).
G Turn on the balance: Press e
> Sartorius logo is displayed
G Select the Weighing in percent application in the Setup menu: Press o
G Select
Application parameters: press the q soft key 2+,
then the
O soft key once
G Select
Application 1 (basic settings): press the O soft key
G Select
Percent weigh.: repeatedly press the Q or q soft key
G Confirm
Percent weigh.: press the O soft key
Percent weigh. Weight storage ο Display accuracy
accuracy +1 decimal place
+2 decimal places
Decimal places None
for readout in pct. 1 decimal place
ο 2 decimal places
3 decimal places
4 decimal places
5 decimal places
6 decimal places
Display calculated ο Residue
value Loss
Ratio1 (DR)
Ratio2 (OR)
ο = factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
G Save settings and exit the Setup menu: press the
oo soft key
Equations
The following equations are used for the corresponding calculations:
Residue (weighing in percent) = Current weight / 100% weight + 100%
Loss (percent-DIFF:) = (Current weight – 100% weight) / 100% weight + 100%
Ratio1 (percent-Ratio 1:) = (100% weight – current weight) / current weight + 100%
Ratio2 (percent Ratio 2:) = 100% weight / current weight + 100%

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius LA Reference Series and is the answer not in the manual?

Sartorius LA Reference Series Specifications

General IconGeneral
BrandSartorius
ModelLA Reference Series
CategoryScales
LanguageEnglish

Related product manuals