56
Examples
Weighing in Percent with Reference Weight Taken From Weight on Balance
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Weighing in percent
Setup: App: Printout: Application-defined output: Auto print upon initialization: All values
Step Key (or instruction) Display/Output
1. Delete previous setting if necessary c
2. Prepare a container Place the empty container
for the parts on the balance
3. Tare the balance w
4. Place the reference weight Place weight equal
on the balance to reference weight in
(here: 1821.48 g = 100%) the container
5. Initialize the balance
Start soft key
6. Unload the balance Remove reference
sample from the container
7. Determine the percentage Place sample to be
of an unknown weight measured in the container
8. If desired, print percentage r
(here: 98.37%)