Choosing a Suitable Location
In order to ensure that the analyzer
functions properly, select a location that
meets the following requirements:
§ Permissible ambient environment
Temperature:
+15 to +40°C (59°F to 104°F)
Relative humidity:
10 – 85%, non-condensing
Altitude:
0 to 2,000 meters (0 to ~6562 ft)
§ The analyzer is intended for indoor use�
§ Place the analyzer on a rigid, horizontal
surface, preferably free of vibration�
§ Avoid drafts and excessive temperature
fluctuation�
§ Locate the analyzer away from
flammable materials�
Connecting the Moisture Analyzer
to AC Power
This Class I Equipment (grounded
type) analyzer contains a universal
self-adjusting power supply rated to
100-120/200-250V~5/2�5A 50/60 Hz�
Check that this voltage matches your
local line voltage� If your voltage is not
within this rating, do not connect the
power cord to the power supply�
Danger
!
The analyzer may only be oper-
ated using the original power cord
supplied� Please do not use it with
an extension cord� The wall socket
must be grounded� The AC Power
mains supply fluctuations are not
to exceed ±10% of the nominal
supply voltage� The (Over Voltage)
Category II for transient over volt-
ages is EN 61010-1� The pollution
Degree 2 is supplied EN 61010-1�
The analyzer is rated to Normal
Protection (not protected against
harmful ingress of moisture)�
§ Insert the power cord that with
the analyzer into the power input
receptacle on the back of the
heater module�
§ Then plug the power cord into a
wall socket�
§ Turn the analyzer on by moving
the power switch on the back of
the analyzer to the On position�
The analyzer will briefly proceed
through a self-diagnostic routine
and then display the standby
screen� It will automatically enter
a warm-up period as noted on the
screen� Should the analyzer display
any warnings, consult the trouble-
shooting section of this manual,
or call Sartorius�
13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius LMA100P and is the answer not in the manual?

Sartorius LMA100P Specifications

General IconGeneral
BrandSartorius
ModelLMA100P
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals