Time 1 and 2
  Each the Temperature 1 and 2 setting 
can have a set period of time� Time 1 or 
2 is the time during which the sample 
will be dried at each temperature before 
advancing to the next temperature�  
Once the time for the last temperature 
has elapsed, the test will terminate 
unless a slope endpoint has been turned 
On� If Temperature 2 is set, then at the 
end of Time 1, the analyzer will ramp to 
Temperature 2�
  To set or change Time 1 or 2:
§  Press the Time 1 or Time 2 soft key to 
display the Time 1 or Time 2 screen�
§  Use the numeric keys type in the desired 
temperature, then press the Enter key� 
The display will return to the Tempera-
ture 1 or 2 screen�
Slope
  The slope function provides an auto-
matic endpoint for the test� The slope 
function consists of two variables:  
window of time and percent of initial 
weight (%IW) change� During the test, 
the weight change of the sample is 
continuously monitored within a mov-
ing window of time� When the loss of 
weight within the window is less than 
the initial weight percentage set, the 
slope criteria has been met� The final 
weight is taken and calculation is  
performed to end the test� This slope  
is referred to as “actual slope” because 
the endpoint is actually achieved�
35
Time
e
i 
g
h
t 
% of I.W.
indow of time
End of test (Endpoint)
W