Leveling the Moisture Analyzer
In order to function properly, the heater
module should be horizontal� The heater
module has a level vial� You can tell
when the analyzer is level by checking
whether the bubble is centered in the
vial�
§ Check that all four feet are in the fully
inserted position�
§ Observing the level vial, rotate one foot
at a time to level�
Adjust control module so that it does
not rock�
Note
The heater module must be re-leveled
each time the analyzer is moved�
Keypad
Warm-up
To enable the analyzer to function
optimally, plug it in and leave it pow-
ered on for a minimum of 15 minutes�
The analyzer is preprogrammed to alert
the operator that the device is in the
warm-up period� It is best to allow the
analyzer to complete this warm-up
period before testing samples; however,
this may be overridden by pressing the
Enter key�