Target Limits
The Target limit Menu provides a statistical process
control (SPC) feature that can be set to identify when
a tested sample is within or outside the desirable limits
selected� If this feature is activated, a Pass or Fail will
be displayed after each test, indicating the sample is
within or outside the set limits; this will also appear on
the printout�
To turn the Target Limits feature on and to enter limits:
§ Press the 8 key from the Program Menu to display
the Target Limits screen� The current target limits are
shown�
§ Press the On|Off soft key to toggle target limits: On
§ Press the 1 key to set the upper limit�
§ Press the 2 key to set the lower limit�
When all the parameters for target limits have been
changed, press the Arrow key to accept the changes�
The display will return to the Program Menu�
Note only for LMA100P:
The limits are set in the current unit with the
exception of ppm Moisture and ppm Solids�
These are set in percent� Example: If you wish to set a
lower limit of 50 ppm M, enter 0�005�
(ppm / 10,000 = %)
40
1. Upper Limit 100.00% Moisture
2. Lower Limit 0.00% Moisture
Targ et Limits Menu
Default 09/07/2012 02:14 PM
On/off