78
Concentration Mode
  The device is able to determine Total 
Suspended Solids in low solids samples, 
such as for wastewater samples� The 
sample solids are first concentrated 
onto a filter pad placed into a vacuum 
apparatus� During the test mode the 
operator will be prompted to enter the 
number of milliliters of sample that was 
filtered� Results are displayed in mg/l�
Wet wt of filter pad (mg) – dry wt of filter pad (mg)
Entered milliliters of sample / 1000
To select the concentration mode:
§  Press the 9 key from the Program menu 
to display the Mode screen�
§  Press the 3 key for Concentration�  
The display will return to the Program 
menu�
To test samples in the Concentration 
mode:
§  Open the heater hood, place the filter 
pad onto a pan and close the hood�
§  Press the Start|Stop key to begin a test�
§  The analyzer will automatically tare the 
pan and filter pad�
§  Open the hood, remove the filter pad 
and place it into the vacuum apparatus� 
Pour a known volume of sample into 
the vacuum apparatus� When finished, 
remove the sample pad and return it to 
the pan in the heater module and close 
the hood�
§  At the prompt, enter the number  
of milliliters of sample that were put 
through the vacuum apparatus and 
then press the Enter key�
§  The analyzer will enter the test mode� 
When the test is done, the results will be 
reported in mg/l�