67
Transfer Programs
The analyzer has been designed for  
easy uploading or downloading of all 
programs from other Sartorius moisture 
analyzers to the device� The follow-
ing combination of analyzers with the 
appropriate cable is capable of transfer-
ring programs:
Mark 3 to LMA1… (serial to serial,  
null modem cable)
Mark 2 to LMA1… (serial to serial,  
see Accessories on page 81)
  Set the same communication parameters 
on both analyzers:
§  Suggested settings: 9600 baud, 8 bits, 
No parity, and 1 stop bit, CTS Off
  To download programs from device to 
another analyzer:
§  Install the appropriate cable between 
the analyzers�
§  Turn the serial port On in serial port 
menu (page 53)�
§  From the Security Menu screen, press 
the 6 key or highlight Manage program 
storage and press the Enter key to 
display the Manage program storage 
screen�
§  (Mark 2 only—press Setup, Setup 2 
menu, Prg Xmit and External input)
§  Press the 4 key for transfer programs�
§  Press the 1 key for LMA1… to Mark 3 or 
2 key for Mark 2 to LMA1… (Mark 3)� 
After a confirmation screen, the  
display will show the program transfer 
in progress�
§  Check the transfer in the receiving ana-
lyzer under Recall�
  To upload programs to your LMA1… 
from another analyzer, follow instruc-
tions on the downloading analyzer:
§  Install the appropriate cable between 
the analyzers�
§  Turn the serial port On in serial port 
menu (page 53)�
§  (Mark 2 only—press Setup, Setup 2 
menu, Prg Xmit and Output Programs)
§  The LMA1… will automatically  
recognize receipt of the programs�
§  Check the transfer in the receiving  
analyzer under Recall�
1) LMA100P to LMA100P 
2) LMA100P to Mark3
3) LMA100P to Mark2
Program  transfer