EasyManua.ls Logo

Sartorius LMA100P - Window of Time and Endpoint Calculation

Sartorius LMA100P
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To set or change slope:
§ Press the 4 key from the Program Menu to display the
Slope screen� The current initial weight percentage is
shown�
§ Use the numeric keys type in the desired %IW, then
press the Enter key�
§ Note that there are some advanced options indicated by
Time and Calc on the soft keys� To change these fea-
tures, see “Window of Time” or “Calculation Endpoint�”
§ When all the parameters for Slope have been changed,
press the Enter again or the Arrow key to return to the
Program menu�
Window of Time
To set or change window of time:
§ Press the Time soft key to display the Window of time
screen�
§ Use the numeric keys to type in the desired window
of time, then press the Enter key� The display will
return to the Slope screen�
Endpoint Calculation
There are three additional advanced slope settings
called Calc 1, Calc 2 and Calc 3, which are algorithms
that calculate the endpoint prior to achieving the actual
slope� These will provide more rapid results on some
samples; however, the success of the Calc mode varies
from case to case�
To set or change Calc endpoint:
§ Press the Calc soft key to display the Calc menu�
§ Press a number key to select the desired calculation set-
ting� The display will return to the Slope screen�
§ If developing a method, check the correlation of each
Calc mode starting with Calc 1 to the actual endpoint
result�
36
Enter percent of initial weight: 0.05
Range: 0.00 to 9.99%
Enter 0.00 to turn off slope
Time: 1.0 Minute
Calculation: Actual
Slope
Default
| %
Time
New
Calc
0 5/01/2012
02:24 PM

Table of Contents