Table: Program Settings
Screen  Parameter Variables  Explanation
Units  Units  Moisture  Initial wt� – final wt� / Initial wt� + 100
    Solids  Final wt� / Initial wt� + 100
    Volatiles  Initial wt� – final wt� / Initial wt� + 100
    Moisture/dry  Initial wt� – final wt� / Final wt� + 100
    ppm moisture*  % Moisture + 10,000
    ppm solids*  % Solids + 10,000
    mg/l  Final wt� + 1000 / volume in mL
    Weight  Initial wt� – final wt�
    TGA moisture*  See appendix
    TGA solids*  See appendix
  Result display  Decimal place  0�01% or 0�001%
  Factors  Scale factor  Factored result = (result + scale factor) + offset
    Offset  Factored result = (result + scale factor) + offset
Temperature 1  Temperature  Maximum 210°C  Set temperature or Off
  Ramp rate*  Standard  Standard temperature ramp rate
    Rapid  More rapid ramp rate
  Time  10�0 – 99�9 minutes  Set time or Off
Temperature 2  Temperature  Maximum 210°C  Set temperature or Off
  Ramp rate*  Standard  Standard temperature ramp rate
    Rapid  More rapid ramp rate
  Time  20�0 – 99�9 minutes  Set time or Off
Slope  % Initial weight  0�010 – 9�990  Weight variable of endpoint criteria
  Window of time  0�1 – 99�9 minutes  Time variable of endpoint criteria
  Calculation  Actual slope  Slope criteria met for endpoint
    Calc 1  Calculated endpoint
    Calc 2  Calculated endpoint
    Calc 3  Calculated endpoint
Start delay  Start delay  0 - 9 seconds  Delay before starting test
  Mode  Mark 2  Simulates Mark 2 start delay
    LMA100P | LMA110S  Start  delay
Standby temp�  Standby temp�  Maximum 165°C  Set temperature between tests
  Equilibration time*  1 – 60 minutes  Delay after standby temperature met
Ideal weight  Ideal weight  0�1 – 99�0 g  Set ideal sample weight for test
  Percent  1 – 10%  Set tolerance of ideal weight
  Ideal weight  Lock On or Off  Set ideal weight lock strict (On) or Off
Target limit  Target limit  Upper limit  Set upper limit of result tolerance
    Lower limit  Set lower limit of result tolerance
  Turn limits  Off  Turns target limit feature Off
Mode  Mode  Standard  Standard test mode
    Syringe*  Initial weight taken by difference
    Concentration  Corrects a result for sample volume
* =  only for model LMA100P
31