LMA100P: Equilibration Time
  The equilibration time is a time in minutes that the 
standby temperature must be maintained before the 
next test can be started�
  To set or change equilibration time:
§  Press the Equil soft key to display the equilibration 
screen�
§  Use the numeric keys to type in the desired time, then 
press the Enter key� The display will return to the 
Standby temperature screen�
Note
  If equilibration is activated, a test cannot be initiated 
with the Start|Stop key until the set equilibration time 
expires� Trying to run a test beforehand will  
a message to appear in the warning box� The status bar 
will show “Ready to Test” when the equilibration period 
has been met�
Ideal Weight
  The ideal weight is the amount of sample in grams that 
should be used for the test as prompted with  
a bar graph and beeper in the Test mode�
  To set or change the ideal weight:
§  Press the 7 key from the Program Menu to display the 
Ideal Weight screen� The current ideal weight  
is shown�
§  Use the numeric keys to type in the desired ideal weight 
then press the Enter key�
§  Note that there are some advanced options indicated by 
Lock and % on the soft keys� To change these features, 
see “Ideal Weight Percent Limit” and “Ideal Weight 
Lock�”
§  When all the parameters for slope have been changed, 
press the Enter key again or the Arrow key to return to 
the Program Menu�
38
Enter  ideal  weight:    5.0  g 
Range: 0.1  –99.0  g 
Ideal  weight  percent:  10%
Lock:  Off 
Ideal  weight 
Default 
09/07/2012  02:14  PM 
|      g 
Lock 
Clear 
%