PR 5210 Operating Manual SMA Protocol
Sartorius 35
5.3.5 Request Scale to Tare
Command: <LF>T<CR>
Response: Scale attempts to tare itself using the weight on the scale display and reports the tare
status in the <s> and <n> status indicators.
For this function the standstill condition of the scale must be fullfilled. The maximum wait
time for standstill could set with the 'Standstill time-out'. See 4.7.2.9
<LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1 Standard Scale Response Message
5.3.6 Set Scale Tare Weight
Command: <LF>T<xxxxxx.xxx><CR>
Response: Scale attempts to take the <xxxxxx.xxx> data as the tare weight and reports the tare
status in the <s> and <n> status indicators.
<LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1 Standard Scale Response Message
5.3.7 Return Tare Weight
Command: <LF>M<CR>
Response: Scale returns tare weight stored in scale Memory and identifies the weight as tare in the
<n> status indicator.
<LF><s><r><T><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1 Standard Scale Response Message
5.3.8 Clear Scale Tare Weight
Command: <LF>C<CR>
Response: Scale Clears tare weight and reports the tare status in the <n> status indicator. The tare
of scale is reset.
<LF><s><r><G><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1 Standard Scale Response Message
5.3.9 Invoke Scale Diagnostics
Command: <LF>D<CR>
Response: The scale runs scale diagnostics and sends a diagnostic response message with the results
of the tests.
<LF><r><e><c><m><CR>
For detail see chapter 5.4.4 Diagnostics Command Response
5.3.10 About Scale First Line
Command: <LF>A<CR>
Response: The scale will send the first line of the About scale data.
<LF><SMA>:<yyyyyy><CR>
For detail see chapter 5.4.5 About ‘A’ and ‘B’ Command Response
5.3.11 About Scale Scroll
Command: <LF>B<CR>
Response: The scale will send the rest of the ABout the scale data.
<LF><MFG>:<yyyyyy><CR>
For detail see chapter 5.4.5 About ‘A’ and ‘B’ Command Response