Profibus Interface PR 5210 Operating Manual
52 Sartorius
6.2.5 Register 4: Calibration information, error byte
Only reading is allowed
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 EXPO
Byte 1 UNIT
Byte 2 STEP
Byte 3 LASTERROR
EXPO One byte for postion of decimal comma/point (Exponent), content in
decimal representation 0 ... 255.
0 = 0000
1 = 000.0
2 = 00.00
3 = 0.000
UNIT One byte for the weight unit, content in decimal representation: 0 ...255
1 = mg milligram 2 = g gram
3 = kg kilogram 4 = t ton
5 = lb pound 6 = l liter
STEP One byte for the stepwidth, content in decimal representation: 0 ...255
1 = stepwidth '1' 2 = stepwidth '2' 5 = stepwidth '5'
10 = stepwidth '10' 20 = stepwidth '20' 50 = stepwidth '50'
LASTERROR Last Error Byte, see also bit CmdError ,number of 'last error':
31 = stand still not reached (e.g. at taring, calibrating)
33 = negative weight at taring and W&M on
35 = weight exceeds allowed range
40 = CAL switch 'locked'
41 = transmitter not in calibration mode
42 = calibration active, transmitter is in calibration mode
46 = tare active (can occur at start calibration)
47 = zero set not executed, weight is not within zero set range
50 = invalid step width
51 = not enough counts/d
53 = FSD < calibration weight
55 = arithmetic overflow
57 = entered unit does not comply with FSD weight unit
58 = span above maximum
59 = Fullscale cannot be divided by stepwidth
30 = weight smaller than deadload
102, 103 = EAROM error (command SaveProcess, register 2)
104 = wrong access code
106 = baudrate of remote display cannot be altered
107 = no standstill at Getfixtare
108 = parameter not valid (at entering via PLC)