Profibus Interface PR 5210 Operating Manual
50 Sartorius
6.2 Profibus Register
6.2.1 Register 0: IO-Status bits for reading
(dynamic status), only reading is allowed
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
byte 0 Input 3 Input 2 Input 1
byte 1 Output 3 Output 2 Output 1
byte 2 Limit 3 Limit 2 Limit 1
byte 3
6.2.2 Register 1: Scale status
(dynamic status), only reading is allowed
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
byte 0 DIM STND INZSR CZERO BELOW0 OVL >FSD ADUERR
byte 1 E1 E3 E7
byte 2 PowerFail ActionActiv CmdError
byte 3 TareActiv CalChanged TestActiv
Byte 0 corresponds to byte 7 in the output area, for weight errors see table in chapter 4.11.1 Analog part
weight status.
ADUERR Errror in analog conversion/ load cell circuitry (OR-function of bits E1,E3,E7)
>FSD gross value larger FSD (german SKE), scale range exceeded
OVL scale overloaded, gross > FSD + Overload, Error 2
BELOW0 Gross weight negative (< 0-1/4 d)
CZERO Center of zero, weight within 1/4d range
INZSR Gross weight is within zeroset range
STND Scale is in stand still
DIM Gross weight has exceeded the scale range (0-1/4 d > weight > FSD+overload), (OR-funktion
of bits BELOW0, OVL).
E7 Measuring signal is negative (inverse conversion), Error 7
E6 Sense voltage missing / too low, Error 6
E3 Measuring signal is > 36mV (no end of conversion), Error 3
E1 Arithmetic error (overflow), Error 1
CmdError Error during execution (cmdError), e.g. action 'taring' is not executed as there is no standstill.
In LastError (register 4) the error is specified. With ResetError register 89 / 89+128 or
register 121 the bit is reset again.
ActionActiv Action will be executed, is still processed
PowerFail Power failure, will be set at each power-on. With bit ResetPWF register 85 / 85+128 or
register 117) 'Power failure reset' the bit PowerFail is reset.
TestActiv Analog test is aktive
CalChanged Calibration mode is active or calibration data have been changed.
FSD has to be read again to reset this bit.
TareActiv Transmitter has beeen tared