SMA Protocol PR 5210 Operating Manual
36 Sartorius
5.3.12 Scale Information
Command: <LF>I<CR>
Response: The scale will send the first line of the scale Information data.
<LF><SMA>:<yyyyyy><CR>
For detail see chapter 5.4.6 Scale Information ‘I’ and ‘N’ Command Response
5.3.13 Scale Information Scroll
Command: <LF>N<CR>
Response: The scale will send the rest of the scale INformation data.
<LF><TYP>:<yyyyyy><CR>
For detail see chapter 5.4.6 Scale Information ‘I’ and ‘N’ Command Response
5.3.14 Abort Command
Command: <ESC>
Response: This is the only command that the scale receives which does not follow the <LF>c<CR>
protocol and does not have a response. The <ESC> character can be detected at any time
and any command is aborted.
5.3.15 Repeat Displayed Weight Continuously
This is a command with Unsolicited Response. It defines a host command that is not considered strictly
command/ response. This is because the scale will respond continuously as commanded to do so by the
host.
Command: <LF>R<CR>
Response: Scale Repeats weight and status information continuously until another command is
received.
<LF><s><r><n><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1 Standard Scale Response Message
Depending on the used baud rate, the following approximately repetition times of response messages
are possible:
19200 Bd Ö 100ms
9600 Bd Ö 110ms
4800 Bd Ö 170ms