Setup and Configuration PR 5210 Operating Manual
26 Sartorius
4.7.2.13 Zero track repeat
With the scale in standstill condition and the gross weight within the zero set range (Zeroset range),
automatic zero tracking is stepwisely done (Zerotrack step) at defined intervals (Zerotrack repeat).
Entry is in multiples of measuring time, permissible range 0 to 100, default = 0 (Automatic zero
tracking = off)
• Switching off the automatic zero tracking is by setting Zerotrack repeat to = 0.
4.8 Parameter
4.8.1 Analog output
For PR 5210/00 und PR 5210/10 only.
4.8.1.1 Analog mode
The following selections are possible:
off
Analog output is not used
transparent
Analog output is controlled via PLC (e.g. to set the mixer speed)
gross
Gross weight value is linked to the output
net
Net weight value is linked to the output (if not tared : gross)