PR 5210 Operating Manual Profibus Interface
Sartorius 53
6.2.6 Register 5: Transmitter type and version
Only reading is allowed
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 TYPE MSB
Byte 1 TYPE LSB
Byte 2 MAINVERSION
Byte 3 SUBVERSION
e.g. 5210 Rel 1.23 = 52100123
hex
6.2.7 Register 6: Board number
Only reading is allowed
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 Board number MSB
Byte 1 ""
Byte 2 ""
Byte 3 Board number LSB
e.g. 148388723 = 08D83B73
hex
6.2.8 Register 7: (Reserved)
6.2.9 Register 8 ...14: Weight data
Only reading is allowed
Gross, net, tare
Are stored as DINT-fixed point. The real datavalue is derived from DINT and EXPO as follows:
Value
Real
=Readout
DINT
*10
(-EXPO
Register 8 Actual gross value
Register 9 Actual net value, if tared, else gross
Register10 Actual tare value, if tared, else 0
Register11 Reserved
Register12 Reserved
Register13 Reserved
Register14 Full scale deflection FSD
Register15 Reserved (free)
6.2.10 Register 20 and 21: Parameter channel (read/write)
Register 20 Parameter value
Register 21 Parameter index