CAUTION
During this operation there is X-ray output.
All the safety measures relevant to radioprotection must be observed.
If necessary, “ X-Mind
DC ” can be calibrated.
The relevant procedure is as follows:
a. Switch off the timer
b. Switch on the timer and keep the mA button depressed
c. The display shows the “tube” message
d. Select X-RAY with the
button to calibrate
e. Take the exposure:
Press and keep the “RX”
button depressed until the beeping stops
stops the yellow “X-RAY OUTPUT SIGNAL” LED goes off
f. Once the exposure has been taken, if the display does not show “ERRORS”, the calibration
has been successfully completed, otherwise (refer to §14 “ERROR MESSAGES”)
13. CALIBRATION