Operating connected seca measuring devices • 69
English
17-10-01-266-002a/07-2018B
6.8 Performing final work
Once you have completed the retrofitting of the measuring devices, perform
the following steps:
Ensure that the device is positioned on the ground flat and stable.
Ensure that no cables or other parts touch the weighing platform.
Perform a function check of the device as described in the respective
instructions for use.
Remove all tools and materials.
Clean the floor and other installation areas.
Clean and disinfect the measuring devices as described in the respective
instructions for use.
Clean and disinfect your hands.
You have the following options to continue:
Add a seca measuring device in the
secaconnect 103 software:
Adding a seca measuring device
Change the configuration of a seca measuring device in the
secaconnect 103 software: Changing a setting of a seca measuring
device
7. OPERATING CONNECTED SECA MEASURING DEVICES
Functional limitations If you connect seca measuring devices to the secaconnect 103, a number of
functions of the seca measuring devices may not be used or only used with
limitations. Otherwise, the following faults can occur:
Malfunction of individual devices
Transmission of invalid measured values
Incorrect allocation of measured values to patient files
Observe the instructions for use of the seca measuring devices.
Inform your users about these limitations. To do so, use the quick
reference “Measurement procedure” Annex: Quick reference:
Measurement procedure.
Functional limitations
Workflow for measurement procedure
Function Device type How to use
hold All Do not use
bmi
Column scale
Measuring stations
Multi-functional scales
Do not use
send/print
a
All seca 360° devices Do not use
2 in 1
Column scale
Measuring stations
Press the button after each
measurement to delete 2 in 1
value from device memory
Deactivate ultrasound length
measurement (for example,
models
seca 287, seca 286)
clear Measuring stations
Press button
before each
measurement to delete old
length measurement value from
device memory