76 • secaconnect 103/seca 452
17-10-01-266-002a/07-2018B
12.TECHNICAL DATA
12.1 seca connect 103 software
seca TestModule: Barcodes You can use the following barcodes to perform system tests with the
seca TestModule integration module. Please contact seca Service if you
have any questions.
Er:8:93
No connection to EMR system
Data is not transmitted or
transmission to EMR system is
defective
- Er:8:94
seca measuring device
blocked, for example, while a
barcode is being scanned
End current action
Restart measurement procedure
-
No connection to the network
(Ethernet/WiFi)
seca measuring device appears in
the Device list as Offline
Visual error message
seca 452
Display
seca measuring
device
Cause Remedy
seca connect 103 software
seca 452 interface module
seca TestModule: Barcodes
Integration module Cerner VitalsLink: Parameter
Barcode for function test with
seca TestModule
Function
Barcode
User ID
Patient ID
Confirm