70 • secaconnect 103/seca 452
17-10-01-266-002a/07-2018B
Workflow for measurement
procedure
To record measured values with seca measuring devices and to send to an
EMR system, the procedure described in this section must be adhered to.
Otherwise, the following faults can occur:
Malfunction of individual devices
Transmission of invalid measured values
Incorrect allocation of measured values to patient files
Observe the instructions for use of the seca measuring devices.
Inform your users about this procedure. To do so, use the quick reference
“Measurement procedure” Annex: Quick reference: Measurement pro-
cedure
tare
Baby scales
Multi-functional scales
Press the button after each
measurement to delete tare value
from device memory
BMIF Baby scales
Deactivate before measurement
is confirmed on seca measuring
device
a. The send/print button on the head slide of the seca 285/seca 284 can be
used as usual.
Function Device type How to use
seca 452 interface
module
seca measuring
device
Patient User
1.
Ensure that the device is switched on
Ensure that Power LED and Network LED
are green
2.
Scan patient/user IDs (according to
Device
settings
)
3.
Ask patient to step onto the scale
4.
Wait until measured value is shown
continuously