From the Service menu press to access the Linear Coefficients screen.
NOTES:
Changing linear coefficients can compromise the analyzer’s accuracy and ability to
function properly.
To reset Factory Coefficients see Reset Factory Settings.
Select the corresponding function key for the range you wish to edit. In the example, was
pressed to allow the Range 1 Linear Coefficient to be changed.
NOTE: The Linear Coefficients function allows the operator to optimize linearity by
inputting up to five coefficients for each range to generate up to a fourth-order curve.