SERVOPRO NOx Analyzers Page
MAXIMUM VERIFYING ERROR
→
→
→
→
→
To access the Maximum Verifying Error screen, press from the Deviation Limits
menu.
To set the allowable tolerances for different ranges, use the up or down arrow
to move the highlight to the field you intend to change.
Press Enter to open the field to change the value in %.
Press Enter again to close the field.
Press to save your changes.
NOTE: Maximum Verifying Error is the allowable tolerance during the Verifying step
of sequenced calibration.