ANALOG HOLD ON CAL
To access the Analog HoldOon Cal menu, press from the Calibration Setup menu.
From the Analog Hold On Cal menu, press to turn Analog Hold On
This will hold the analog outputs at the last measured value.
Once pressed it will return to the Calibration Setup menu.
The Calibration Setup menu shows the current setting at the bottom-right corner
of the screen after the (. . .).
Example: Analog Hold on Cal . . . On.
From the Analog Hold On Cal menu, press to turn Analog Hold Off.
This will allow the analog outputs to be live and reporting out values during
calibration.
Once pressed it will return to the Calibration Setup menu.
NOTE: Analog Hold on Cal will hold the analog outputs to the last measured value
during calibration.