SERVOPRO NOx Analyzers Page
17
SITE SELECTION AND MOUNTING
CAUTION: The following precautions must be carefully observed:
1. Select a site free from direct sunlight, radiation from a high-temperature
surface, or abrupt temperature variations.
2. This analyzer is not suitable for installation outdoors.
3. Select a site where the air is clean. Avoid exposing the instrument to corrosive
or combustible gases.
4. The instrument must not be subject to severe vibration. If severe vibration is
present, use isolation mounts.
5. The instrument is designed for rack mounting. Optional rack-mount slides are
available.
6. Do not install the SERVOPRO NOx analyzer near equipment that emits
electromagnetic interference (EMI).
The Power On/Off switch is accessible from the rear of the analyzer only.
DO NOT mount the analyzer in a manner that leaves the Power On/Off
switch inaccessible.
NOTE: A front and rear supporting brace or equivalent is required if the optional
rack mount slides were not purchased.