MAXIMUM CALIBRATION ERROR
Each range has its own set of Absolute and Relative tolerances. The deviations must be
inside these tolerances to accept a calibration. To access the Maximum Calibration
Error screen, press from the Deviation Limits menu.
To navigate between fields, use the up or down arrow to move the
highlight to the field you intend to change.
Press Enter to open the field to change the allowable tolerance in %.
Press Enter again to close the field.
Press to save your changes.
Absolute Deviation is used to compare the factory-set calibration to the current
calibration.
Relative Deviation compares the current calibration to the previous calibration.
NOTE: Maximum Calibration Error is used by the operator to define the maximum
acceptable tolerances for Absolute and Relative deviations.