SERVOPRO NOx Analyzers Page
64
CAL GAS CONCENTRATIONS
or
To access the Cal Gas Concentrations screen (shown above) press from the
Calibration Menu.
The Cal Gas Concentrations screen displays the range identification, the changeable span
gas value and the full-scale value set for that range.
Using the Up/Down arrows move the highlighted field to the span gas value
you wish to change (for example, 2.850 above). Press the Enter button to open the
span gas value field and change the value to match the span gas being supplied to the
analyzer. Press the Enter button again to close the span gas value field. Press
to save the changes.
NOTE: The Cal Gas Concentrations screen allows operators to change calibration gas
values for multiple ranges on one screen.