INITIATE SEQUENCED CHECK
→
→
→
→
From the Auto Calibration Menu screen, pressing initiates a sequenced calibration
check. Rather than initiating a calibration, it checks the validity of your most current
calibration.
A sequenced calibration check performs all of the steps of a sequenced calibration with
the exception of the zero and span calibrations. It does not set new offsets, gains or any
alarms.
NOTE:
Initiate Sequenced Check is a useful tool for setting up Auto Calibration.