SERVOPRO NOx Analyzers Page
170
11 = Valves
AFDA: Auto Calibration Times and Purge Time
Command Response Description
_AFDA_K0_SATK
_AFDA_s_z_y_x_w
Auto calibration times in seconds:
z: Purge time
y: Verify time
x: Purge after time
w: Calibrate time
v: Total time
_AFDA_K0_SSPL _AFDA_s_z Purge time will be responded
AFGR: Default Factory Polynomial Coefficients
Command Response Description
_AFGR_K0_Mn _AFGR_s_a0_a1_a2_a3_a4 Factory coefficients for range n
_AFGR_K1_M1 _AFGR_s_a0_a1_a2_a3_a4 K1 O
2
option
AGRD: Polynomial Coefficients
Command Response Description
_AGRD_K0_Mn _AGRD_s_a0_a1_a2_a3_a4 Polynomial coefficients of channel m Range n
_AGRD_K1_M1 _AGRD_a0_a1_a2_a3_a4 O
2
coefficients with option
AGRW: Max Absolute / Relative Deviation Limits
Command Response Description
_AGRW_K0_Mn _AGRW_s_z.z_y.y z: Absolute
y: Relative for range n
_AGRW_K1 _AGRW_s_z.z_y.y K1 O
2
option
AKAK: Calibration Gas Concentrations