SERVOPRO NOx Analyzers Page
59
range must be calibrated separately). This also applies if the analyzer is in auto
range.
This automated calibration is triggered manually and not by the analyzer’s clock
or via remote signal. A fully automated sequenced calibration can be preset to
include the desired interval for recurring analyzer-initiated calibrations. This
requires additional setup. Automatic calibration of multiple ranges is also
possible. See
Calibration Setup.
If a sequenced calibration was unintentionally started, pressing the back
button before the Zero step is completed will cancel the calibration.