SERVOPRO NOx Analyzers Page
90
RESET OFFSETS AND GAINS
Pressing from the Reset Factory Settings menu will prompt the operator to confirm
resetting the Linear Coefficients for all ranges. Pressing (Yes) from this screen
resets the Offset and Gain factors to factory default settings (0 and 1 respectively) and
brings you to this confirmation screen:
Offset and Gain factors are created when the analyzer is zeroed and spanned.
If the Offsets and Gains are reset, the analyzer must be zeroed and spanned again
before use.
All recorded deviations will be set to zero.