WARNING
When executing [Start Leak Check] in the "Initialize" dialog box or [Instrument]-[Gas Leak Check] in the
menu bar, an automatic gas leak check is performed for about 11 minutes.
If the instrument does not detect any fuel gas leakage, the following message is displayed on the screen
after about 11 minutes, after which the flame can be ignited.
If the instrument detects any fuel gas leakage, the following message is displayed on the screen at that time.
In this case, please close the stop valve of the fuel gas immediately.
Once completed, disconnect the communication between PC and the instrument, and turn the power supply
off. Be sure to contact your service representative to check and repair the instrument.
Action in the Case of Fuel Gas Leakage