Page 168 HydroRanger 200 – INSTRUCTION MANUAL 7ML19985FC03
mmmmm
Parameters
P603 Maximum Head [6 relay model]
The level value associated with Maximum Flow, in Units (P005).
This represents the highest head level supported by the PMD and works in conjunction
with Maximum Flow (P604) to define the highest point in the exponential curve. Use it
when the Primary Measuring Device (PMD) requires a maximum head and flow
reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume, and
Universal breakpoints.
P604 Maximum Flow [6 relay model]
The maximum flowrate associated with Maximum Head (P603).
This represents the flow at the highest head level supported by the PMD. and works in
conjunction with Maximum Head (P603) to define the highest point in the exponential
curve. Use it when the Primary Measuring Device (PMD) requires a maximum head and
flow reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume,
and Universal breakpoints.
Also use this parameter with Time Units (P606) to define the flowrate units. The limitation
of four digits is for the LCD only, and the flowrate value is available with greater precision
through communications.
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: -999 to 9999
Preset: Span (P007) value
Altered By
P005 Units
P600 Primary Measuring Device
Related
P604 Maximum Flow
P605 Zero Head
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: -999 to 9999
Preset: 1000
Altered By P600 Primary Measuring Device
Related
P603 Maximum Head
P606 Time Units
P925 Flow Measurement