Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing basic unit components 13
Page 13 of 64
Medical Solutions
Fig. 2: Remove cable cover
Fig. 3: Remove cable cover - different view
Removing the S2 & S3 switches from the grid and grid holders
1. Mark the position where the front and rear carrier brackets (holders) of the magnifica-
tion table locator (or grid position) micro switches S2 and S3 are fastened to the object
table support plate. (Such markings may already have been made at the factory).
These micro switches are on the left side of the plate and they can be adjusted only with
the magnification table. It is not possible to adjust them if the customer has no such ta-
ble. Draw lines around both carrier brackets of these switches so they can be fastened
back in the same position. (See (1/Fig.4/p.14)).