42 Collimator and D814
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 42 of 64
Medical Solutions
7-
6Collimator and D814
Replacing the collimator and D814 0
General 0
These instructions describe the work steps for replacing the collimator and D814.
Required documents 0
Software System Stand, SPB7-250.816.03....
System Start-up with WSH AWS, SPB7-250.815... for systems with AWS or
Installation Instructions, SPB7-250.812... for systems with DROC
QC Manuals
Tools 0
Standard tool kit
Service PC
Disassembling the collimator 0
WARNING
Before you begin working on equipment, it is very important that
you disconnect it from the power supply at the main circuit break-
er.
¹ Before removing or inserting any of the printed circuit
boards, switch off the equipment.
WARNING
If the system is only switched off at the control panel, the line volt-
age will still be present at the generator line connection (see wiring
diagram).
¹ A life-threatening hazard of electric shock exists.