Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Generator 57
Page 57 of 64
Medical Solutions
9-
8Generat or
Replacing generator modules 0
General 0
The generator is located on the back of the stand. It consists of several modules con-
nected by cables. These modules have been developed in such a way that no electrical
settings are necessary in case of a replacement. The modules are mounted on the gener-
ator frame with the help of metal rails. They are each fastened by 2 screws located on the
left side.
Figure 1 shows the generator modules.
Fig. 35: Generator
High-voltage tank
High-voltage inverter
Tube power supply
AC inverter