58 Generator
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 58 of 64
Medical Solutions
• Mains input converter
Required documents 0
Software System Stand, SPB7-250.816.03...
System Start-up with WH AWS, SPB-7.250.815...for systems with AWS or Installation
Instructions and Start-up, SPB7-250.812.04.. for systems with DROC.
QC Manuals
Tools 0
Standard tool kit
Service PC
Oscilloscope
WARNING
It is very important to begin any work on the equipment
by disconnecting it from the power supply via the main circuit
breaker.
¹ Before removing or inserting any of the modules, switch
off the equipment.
Replacing a module 0
• Turn off the system and disconnect it from the power supply.
• Remove the covers from the back of the Mammomat stand.
• Disconnect corresponding plug-in connections of the module.
• Remove the mounting screws (2x screws located on the left side).
• Shift the module to the left and remove it.
• Insert the new module und tighten the screws.
• Reestablish the plug-in connections.
• Switch on the system.
Tube power supply
The frequency of the rotating anode has to be checked after the tube power supply mod-
ule is replaced. To do so, load the stand service software by selecting Main menu->Nor-
mal mode->Generator data.