32 Compression Unit
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 32 of 64
Medical Solutions
Fig. 11: Distance
Calibrating the compression unit 0
Connect the service PC according to Documentation Software System Stand,
SPB7-250.816.03…
Start the service program.
In the main menu, under Configuration -> Compression, select Optimized compres-
sion. If you find the number/value 5137 in any of the (two) OPCOMP data fields shown in
the appearing mask, change it to some other arbitrary number. If the value in those data
fields differs from 5137, press F2 (save).
In the main menu, under Configuration ->Compression, select Optimized compres-
sion dr and repeat the procedure described above.
Prior to calibration, the following values must be entered in the system as default values.
Force = 0 kp 0066
Force = 20 kp 02D8
Preset force 3 kp 06
Preset force 20 kp FF
Thickness bottom 0093
Thickness bottom 0
Thickness top 037E
Thickness top 180